Product Description
For the detection of hydrogen sulfide in liquefied petroleum (LP) gases. The sensitivity of the test is about 4 mg/m3 (0.15 to 0.2 grain of hydrogen sulfide per 100 ft3) of gas
Vaporized LP gas is passed over moist lead acetate paper under controlled conditions. Hydrogen sulfide reacts with lead acetate to form lead sulfide which produces a coloration on the paper varying from yellow to black, depending upon the amount of hydrogen sulfide present. Liquefied petroleum gases and their products of combustion must not be unduly corrosive to the materials with which they come in contact. The potential personnel exposure hazards of H2S also make the detection and measurement of hydrogen sulfide important, even in low concentrations. In addition, in some cases the odor of the gases must not be objectionable.
SPECIFICATIONS:
~ Temperature Range: +5 to 150°C
~ Temperature Control Stability: ± 0.05°C
~ Temperature Readout Accuracy: ± 0.5°C
~ Bath Capacity: 28L (7.3gal)
~ LPG Cylinder Capacity: 500mL
~ Flow Meter Range: 2 - 3 L/min
~ Thermometer Range: -5 to +110°C
The following link provides additional information from the manufacturer's brochure. Click to open the PDF in a new window or right-click to download it:
Koehler K48200 / K48290 Hydrogen Sulfide Apparatus or copy the following into your browser address window: labequip.ca/stock/pictures/n1806.pdf
Made in USA.
ACCESSORIES:
COMES COMPLETE WITH:
KD2420 Glassware Kit
Comprised of:
Glass Cylinder
Rubber Stopper (2)
Watch Glass
Glass Rod / Hook
K27853 LPG Sampling Cylinder, 500mL
OPTIONAL ACCESSORIES AVAILABLE AT ADDITIONAL COST:
K48200-1 LPG Flowmeter Assembly, Range 2 - 3 L/min
K48200-2 Lead Acetate Test Paper, 7mm Width, 5m Roll
250-000-15C ASTM 15C Thermometer