Product Description
Rust Preventing Characteristics Oil Bath determines a lubricant's ability to prevent rusting of metal surfaces. Suitable for steam turbine oils, gear oils, hydraulic oils and other types of inhibited mineral oils. A steel test specimen is immersed in a heated mixture of sample oil and water which is stirred continuously during the test. After the test period the specimen is examined for rusting. The standard (ASTM D665) method uses a vertical specimen; the `horizontal disc method' (ASTM D3603) uses both horizontal and vertical test surfaces.
~ Conforms to ASTM D665, D3603 and NACE TM-01-72* specifications
~ Accommodates six sample beakers
~ Microprocessor programmable high accuracy temperature control
Constant temperature bath with stirrers for rust preventing characteristics tests. Stirs sample-water mixtures at 1000rpm and controls temperature with ± 0.5°C (± 1°F) stability. Immerses test beakers at the proper depth per ASTM specifications.
Microprocessor PID control provides quick temperature stabilization without overshoot, and the bath is protected by an overtemperature control circuit that interrupts power should bath temperature exceed a programmed cut-off point. Dual LED displays provide actual and setpoint temperature values in °C/°F format. Communications software (RS232, etc.), ramp-to-set and other enhanced features are available as extra cost options. Contact Labequip, your authorized Koehler representative for information.
Stainless steel stirrer paddles are driven by a ball bearing type motor through an improved pulley drive-roller bearing arrangement. Paddles can be raised and lowered for placement of sample beakers in the bath. Includes test specimens, holders and beaker covers for ASTM D665 or D3603 testing (see specifications and ordering information). Stainless steel bath includes perforated support shelf for beakers and two-position cover plate that adjusts for either ASTM D665 or D3603 testing. Long-lasting polyester drive belt improves reliability. Drive train components are protected by a removable steel guard. All exterior surfaces have stainless steel or chemical resistant polyurethane enamel finishes.
*To order this equipment for the NACE TM-01-72 test, please see K30160NACE.
**Accessories for these test methods are available upon request.
Testing Capacity: Six (6) 400mL sample beakers
Maximum Temperature: 104°C (220°F)
Temperature Control Stability: ±0.5°C (±1°F)
Drive Motor: 1/12hp induction motor
Bath Medium: 11 gal (41.6L) white technical oil
ASTM D665 Models (K30160, K30165, K30166) is the model priced here. Please request other prices for ASTM D3603 Models (K30161, K30167, K30168)
The following link provides additional information from the manufacturer's brochure. Click to open the PDF in a new window or right-click to download it:
Koehler K30160 / K30165 Rust Preventing Characteristics Bath or copy the following into your browser address window: labequip.ca/stock/pictures/n1172.pdf
Made in USA.
ACCESSORIES:
FOR ASTM D665 - COMES COMPLETE WITH:
Steel Test Specimens (6)
Type 2 Plastic Specimen Holders (6)
Plastic Beaker Covers (6)
FOR ASTM D3603 - COMES COMPLETE WITH:
Horizontal Disc Test Assembly (6) consisting of: plastic beaker cover, horizontal test specimen, vertical test specimen, fluorocarbon washer, plastic cap, stainless steel support rods and hardware
OPTIONAL ACCESSORIES AVAIABLE AT ADDITIONAL COST:
- 332-002-006 Test Beaker, 400mL for ASTM D665 & D3603
- 250-000-09F ASTM 9F Thermometer; Range: 20 to 230°F
- 250-000-09C ASTM 9C Thermometer; Range: -5 to +110°C
- K30130 Chuck for polishing test specimens - Includes locknut and shaft for mounting on accessory drive motor
- K30150 Drive Motor - Drives K30130 Chuck. Mounted on base
- 380-150-002 Aluminum Oxide Cloth, 150-grit for preliminary grinding of test specimens Pack of 50
- 380-240-002 Aluminum Oxide Cloth, 240-grit for final polishing of test specimens Pack of 50
- K30140 Auxiliary Stirrer Blade - Attaches to stirrer shaft - for testing heavier than water samples - ASTM D665. Procedure C.
Test Specimens:
- K30110 Steel Test Specimen for ASTM D665 Machined to ASTM specifications. Without Holder
- K30100 Test Specimen with Type 2 Plastic Holder for ASTM D665
- K30119 Test Specimen with Type 1 Plastic Holder for ASTM D665
- K30101 Test Specimen with Type 2 PTFE Holder
- K30810 Horizontal Test Specimen for ASTM D3603
- K30820 Vertical Test Specimen for ASTM D3603
- K30800 Horizontal Disc Rust Test Assembly for
ASTM D3603. Includes polycarbonate beaker cover, two stainless steel support rods, disc carrier and one each horizontal and vertical test specimens