Product Description
The HI98193 is a rugged, portable dissolved oxygen (DO) meter designed for demanding applications. This professional, waterproof meter complies with IP67 standards and measures DO, barometric pressure, BOD and temperature. The HI98193 is supplied complete with all accessories to perform a DO measurement packaged into a durable carrying case.
For the measurement of DO, the HI98193 is supplied with the HI764073 polarographic dissolved oxygen probe. This probe uses preformed PTFE membrane caps that are field replaceable. The probe connects to the meter with a unique quick connect waterproof DIN connector allowing for a secure attachment without a threaded connect that is also easy to remove.
The HI98193 portable DO meter has extended ranges of up to 50 ppm and 600% saturation. When measuring dissolved oxygen, compensations for pressure, temperature and salinity are essential for an accurate reading.
Pressure compensation can be done automatically with the meter's built-in barometer or by manual entry. Calibration of the built-in barometer is done to one point by entering the value of a reference barometer. Pressure is displayed in user-selectable units including mmHg, inHg, atm, psi, kPa, and mbar.
The HI764073 DO probe has a built-in temperature sensor for automatic temperature compensation. The temperature sensor can be calibrated to one or two points. Manual entry of salinity values allows for the salinity compensation of dissolved oxygen readings in saline waters.
Other features include built-in methods and calculations for the measurement of Biological Oxygen Demand (BOD), Oxygen Uptake Rate (OUR) and Specific Oxygen Uptake Rate (SOUR).
The HI98193's log on-demand feature allows users to store up to 400 readings. This data can then be transferred to a PC with the HI920015 USB cable and HI92000 software.
Features/Benefits:
Ergonomic, rugged, waterproof (IP67) design
Display dissolved oxygen in % saturation or mg/L
Built in Barometer
Automatic barometric pressure compensation with 1 point calibration
Displays pressure in user-selectable units (mmHg, inHg, atm, psi, kPa, mbar)
Built-in temperature sensor
Automatic temperature compensation with one or two point temperature calibration
Displays temperature in °C or °F
Salinity compensation
Manual salinity compensation with a selectable salinity value
Built-in calculations
Determination of Biochemical Oxygen Demand (BOD), Oxygen Uptake Rate (OUR) and Specific Oxygen Uptake Rate (SOUR)
Polarization
Automatic polarization of polarographic probe at startup to ensure that the probe is ready to use when measurement screen is displayed
Membrane caps
Ready-to-use field replaceable PTFE membrane caps
Log-on-demand
Store measurement data at the press of a button
GLP features
GLP data provides calibration data including date, time, pressure, calibrated value, temperature and salinity value of the last calibration
Contextual help at the press of a button
Backlit, Graphic LCD
SPECIFICATIONS:
DO Range: 0.00 to 50.00 ppm (mg/L); 0.0 to 600.0% saturation
DO Resolution: 0.01 ppm (mg/L); 0.1% saturation
DO Accuracy (@25ºC/77ºF): ±1.5% of reading ±1 digit
DO Calibration: automatic one or two point at 100% saturation (8.26 mg/L) and 0% saturation (0 mg/L).
Atmospheric Pressure Range: 450 to 850 mmHg
Atmospheric Pressure Resolution: 1 mmHg
Atmospheric Pressure Accuracy: ± 3 mmHg within ±15% from the calibration point
Atmospheric Pressure Calibration: one point at any in range pressure value
Temperature Range: -20.0 to 120.0°C (-4.0 to 248.0°F)
Temperature Resolution: 0.1°C (0.1°F)
Temperature Accuracy (@25ºC/77ºF): ±0.2°C; ±0.4°F (excluding probe error)
Temperature Calibration: one or two point at any in range temperature value
Temperature Compensation: automatic from 0 to 50°C (32 to 122°F)
Measurement Modes: direct DO; BOD (biochemical oxygen demand); OUR (oxygen uptake rate); SOUR (specific oxygen uptake rate)
Barometric Compensation: automatic from 450 to 850 mmHg
Salinity Compensation: automatic from 0 to 70 g/L
Electrode/Probe: HI764073 polarographic DO probe with protective sleeve, internal temperature sensor, DIN connector and 4m (13') cable (included)
Log-on-demand: up to 400 samples
PC Connectivity: opto-isolated USB with optional HI 92000 software and micro USB cable
Battery Type/Life: 1.5V AA batteries (4) / approximately 200 hours of continuous use without backlight (50 hours with backlight)
Auto-off: user selectable: 5, 10, 30, 60 min or can be disabled
GLP data is accessible by a fixed dedicated key this data includes date, time, pressure, calibrated value, temperature and salinity value of the last calibration.
The compact, durable HI760193 carrying case is thermoformed to hold all necessary components for taking a field measurement, including the meter and electrode, beakers, buffer solutions, cleaning solutions and storage solutions.
The following link provides additional information from the manufacturer's brochure. Click to open the PDF in a new window or right-click to download it:
Hanna HI 98193 Portable Oxygen/BOD Meter or copy the following into your browser address window: labequip.ca/stock/pictures/n1709.pdf
ACCESSORIES:
COMES COMPLETE WITH:
HI 764073 DO probe
preformed PTFE membrane cap (2)
HI 7041S electrolyte solution (30 mL)
HI 7040 bicomponent zero oxygen solution, 100 mL
plastic beakers (2)
HI 92000 PC software
HI 920015 micro USB cable
1.5V AA batteries (4)
instruction manual
quick start guide
instrument quality certificate
HI 760193 rugged carrying case
OPTIONAL ACCESSORIES AVAIABLE AT ADDITIONAL COST:
HI 98501 Digital thermometer (range: -50.0 to 150.0 °C/-58.0 to 302 °F)
HI 7040L Zero Oxygen Solution
HI 7041S Refilling Electrolyte Solution (30 mL)
HI 764073 Spare probe with 4 meter (13.4') cable
HI 76407A/P Spare membranes (5 pcs.)
HI 92000 Windows ® compatible software application
HI 920015 Micro USB cable